Flow injection unit with merging zone technique was designed for aniline blue determination spectrophotometrically. This unit consists of a valve was designed for dye injection. The scope of the study is to find the optimum conditions :volume of sample, flow rate of carrier, dead volume, dispersion coefficient and reproducibility. They found 117µl, 4.3 ml/min, zero dead volume, 2.62 and high reproducibility for six times respectively. The range of aniline blue concentration by flow injection analysis was 0. 1 -8 ppm and detection limit 0.1 with R 2 = 0.9952 and 0.01 -16 ppm for spectrophotometric method with R 2 =0.9947 with detection limit 0.01 .
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.