The Deltatrac II MBM-200 metabolic monitor appears accurate for low levels of V(O2) and V(CO2) during mechanical ventilation with F(IO2) levels up to 0.390. With increasing F(IO2) to 0.558, the increase in deviation of V(O2) for single measurements can be of clinical relevance for mechanically ventilated infants and children. The increased fluctuation of delivered F(IO2) on higher levels of F(IO2) is likely the cause of the inaccuracy.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.