Recently, Siemens introduced a software-based scatter correction in combination with grid-less digital mammography for dose-reduced breast imaging. In this study, the potential dose reduction when compared with using a grid was calculated. Image quality was evaluated using the CDMAM phantom. Nine hundred and sixty CDMAM measurements were performed with and without grid. The images were analysed with the CDCOM software. The correlation of CDCOM readings of gold thickness thresholds with human readings for grid-less imaging was found to be linear, with slope 2.157 and off-set 0.024 (R(2) = 1), validating the further use of CDCOM. Dose reduction for 21, 32, 45, 60, 75 and 90 mm equivalent breast thickness was found to be 28, 17, 21, 18, 8 and 3 %. Scatter correction did not influence CDCOM readings, but image homogeneity was improved. Grid-less acquisitions, with or without scatter correction, can be used for reducing the dose while maintaining image quality as evaluated from the CDMAM phantom.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.