Purpose:
Although the Cochet-Bonnet esthesiometer (CBE) measures corneal sensitivity, it has heretofore only been tested on the index pulp. Tactile skin sensitivity thresholds are measured with Semmes-Weinstein monofilaments (SWM). This study measured skin sensitivity thresholds in healthy individuals using CBE and SWM, and compared both instruments in territories involved in corneal neurotization.
Methods:
Overall, 27 healthy individuals were tested by a single examiner at 9 territories on the face, neck, forearm, and leg, using 20-thread SWM and CBE with a diameter of 0.12 mm. Both sides were tested. Thresholds were compared for both instruments and between the different territories using Bayesian methods.
Results:
Mean sensitivity levels for SWM ranged from 0.010 to 1.128 g, while mean sensitivity levels for CBE ranged from 0.006 to 0.122 g. Thresholds measured with SWM were significantly higher than with CBE. Both instruments demonstrated higher thresholds in the leg territory than the forearm. However, the forearm presented higher thresholds than the head territories. No significant differences were found between the head territories themselves. Overall, right-side territories exhibited lower thresholds than left-side territories.
Conclusions:
We have reported the first mapping of skin sensitivity thresholds using CBE. Thresholds measured with CBE and SWM were coherent. The use of CBE on the skin is particularly relevant to the field of corneal neurotization.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.