A system is described that can simultaneously record cytosolic Ca2+ concentration ([Ca2+]i), cell length, and either membrane potential or current in single cardiac myocytes loaded with the fluorescent Ca2+ indicator indo-1. Fluorescence is excited by epi-illumination with 3.8-microsecond flashes of 350 +/- 5 nm light from a xenon arc. Indo-1 fluoresence is measured simultaneously in spectral windows of 391-434 nm and 457-507 nm, and the ratio of indo-1 emission in the two bands is computed as a measure of [Ca2+]i for each flash. With cells loaded with the permeant acetoxymethyl ester of indo-1, quantitation of [Ca2+]i is not precise, owing to subcellular compartmentation of indo-1; however, the instrument would allow full quantitation if indo-1 free acid was introduced by microinjection. Simultaneously, cell length is measured on-line from the bright-field image of the cell. Because fluorescence collection is time gated during the brief flash, and red light (650-750 nm) is used for the bright-field image, cell length and [Ca2+]i measurements are obtained simultaneously without cross talk. Membrane potential or current can be recorded simultaneously with indo-1 fluorescence and cell length via standard patch-clamping techniques.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.