Optimal function following elbow replacement surgery is dependent on the accurate replication of the elbow's flexion-extension axis. Currently, position and orientation of the axis are estimated from visual landmarks. In order to develop computer-assisted techniques to more accurately define this axis, a surface-based registration technique employing a hand-held laser scanner was evaluated against a conventional paired-point registration method to determine whether it produced improved alignment of the flexion-extension axis of the elbow. Registration error was 0.8 +/- 0.3 mm for surface-based registration, compared with 1.9 +/- 1.0 mm for the conventional registration method. These results suggest that the implementation of a surface-based registration technique may lead to a more accurate axis determination and improved clinical outcomes following elbow replacement surgery.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.