Intraoperative radiotherapy (IORT) is a specialised subset of radiotherapy, where a high radiation dose is delivered to a surgically exposed tumour bed in order to eradicate any remaining cancer cells. The aim of this study was to examine the dose characteristics of the Zeiss Intrabeam IORT device which provides near-isotropic emission of up to 50 kV X-rays. The EGSnrc Monte Carlo (MC) code system was used to simulate the device and percentage depth dose (PDD) data measured with a soft X-ray parallel-plate ionisation chamber were used for model verification. The model provided energy spectra, isodose curves and mean photon energies. In addition, EBT3 Gafchromic film was used to verify the MC model by examining PDDs and 2D dose distributions for various applicators. The differences between MC model and ionisation chamber measurements were within 3% for most points, with a maximum deviation of ~ 9%. Most of the simulated PDD points were within 5% of the film-measured data, with a maximum deviation of ~ 10%. The mean energy of the bare probe was found to be 21.19 keV. The mean photon energy from applicators ranged from 29.00 to 30.85 keV. Results of this study may be useful for future work on creating a system for treatment planning.
For safe and accurate dose delivery in brachytherapy, associated equipment is subject to commissioning and ongoing quality assurance (QA). Many centres depend on the use of a well-type chamber ('well chamber') for performing brachytherapy dosimetry. Documentation of well chamber commissioning is scarce despite the important role the chamber plays in the whole brachytherapy QA process. An extensive and structured commissioning of the HDR 1000 plus well chamber (Standard Imaging Inc, Middleton WI) for HDR and LDR dosimetry was undertaken at Sir Charles Gairdner Hospital. The methodology and outcomes of this commissioning is documented and presented as a guideline to others involved in brachytherapy. The commissioning tests described include mechanical integrity, leakage current, directional dependence, response, length of uniform response, the influence of insert holders, ion collection efficiency, polarity effect, accuracy of measured air kerma strength (S(K)) or reference air kerma rate (K(R)) and baseline setting (for ongoing constancy checks). For the HDR 1000 plus well chamber, some of the insert holders modify the response curve. The measured sweet length was 2.5 cm which is within 0.5% of that specified by the manufacturer. Correction for polarity was negligible (0.9999) and ion recombination was small (0.9994). Directional dependence was small (less than 0.2%) and leakage current was negligible. The measured K(R) for (192)Ir agreed within 0.11% compared with a second well chamber of similar model and was within 0.5% of that determined via a free-in-air measurement method. Routine constancy checks over a year agreed with the baseline within 0.4%.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.