SIGNIFICANCE
Vision screenings are conducted to detect significant refractive errors, amblyopia, and ocular diseases. Vision screening devices are desired to have high testability, sensitivity, and specificity. Spot has demonstrated high testability, but previous reports suggest that the Spot has low sensitivity for detecting amblyogenic hyperopia and moderate sensitivity for amblyogenic astigmatism.
PURPOSE
This study assessed the concurrent validity of detecting amblyogenic refractive errors by the Spot (v.1.1.50; Welch Allyn Inc., Skaneateles Falls, NY) compared with cycloplegic retinoscopy.
METHODS
A total of 475 subjects (24 to 96 months) were screened by Spot and then received a masked comprehensive examination. Sensitivity and specificity, Bland-Altman plot, receiver operating characteristic area under the curve, and paired t test were evaluated by comparing the results of the Spot (v1.1.50) using the manufacturer referral criteria with the results of the comprehensive examination using the 2013 American Association for Pediatric Ophthalmology and Strabismus criteria.
RESULTS
The Spot (v.1.1.50) referred 107 subjects (22.53%) for the following: 18.73% (89/475) astigmatism, 4.63% (22/475) myopia, 0.42% (2/475) hyperopia, and 2.11% (10/475) anisometropia. The sensitivity and specificity of the Spot vision screener for detecting amblyogenic risk factors were 86.08% (95% confidence interval [CI], 76.45 to 92.84%) and 90.15% (95% CI, 86.78 to 92.90%). Areas under the curve were 0.906 (95% CI, 0.836 to 0.976) for hyperopia, 0.887 (95% CI, 0.803 to 0.972) for spherical equivalent, and 0.914 (95% CI, 0.866 to 0.962) for astigmatism. A modified hyperopia criteria cutoff of greater than +1.06 D improved the sensitivity from 25 to 80% with 90% specificity. The current cutoff criterion, greater than −1.75 D, for astigmatism seemed optimal.
CONCLUSIONS
This study shows that the Spot vision screener accurately detects low spherical refractive errors and astigmatism. Lowering the hyperopia cutoff criteria from the current Spot screener referral criteria improves the sensitivity with desired (high) specificity.
PurposeTo survey paediatric eye care providers to identify current patterns of prescribing for hyperopia.MethodsPaediatric eye care providers were invited, via email, to participate in a survey to evaluate current age‐based refractive error prescribing practices. Questions were designed to determine which factors may influence the survey participant's prescribing pattern (e.g., patient's age, magnitude of hyperopia, patient's symptoms, heterophoria and stereopsis) and if the providers were to prescribe, how much hyperopic correction would they prescribe (e.g., full or partial prescription). The response distributions by profession (optometry and ophthalmology) were compared using the Kolmogorov–Smirnov cumulative distribution function test.ResultsResponses were submitted by 738 participants regarding how they prescribe for their hyperopic patients. Most providers within each profession considered similar clinical factors when prescribing. The percentages of optometrists and ophthalmologists who reported considering the factor often differed significantly. Factors considered similarly by both optometrists and ophthalmologists were the presence of symptoms (98.0%, p = 0.14), presence of astigmatism and/or anisometropia (97.5%, p = 0.06) and the possibility of teasing (8.3%, p = 0.49). A wide range of prescribing was observed within each profession, with some providers reporting that they would prescribe for low levels of hyperopia while others reported that they would never prescribe. When prescribing for bilateral hyperopia in children with age‐normal visual acuity and no manifest deviation or symptoms, the threshold for prescribing decreased with age for both professions, with ophthalmologists typically prescribing 1.5–2 D less than optometrists. The threshold for prescribing also decreased for both optometrists and ophthalmologists when children had associated clinical factors (e.g., esophoria or reduced near visual function). Optometrists and ophthalmologists most commonly prescribed based on cycloplegic refraction, although optometrists most commonly prescribed based on both the manifest and cycloplegic refraction for children ≥7 years.ConclusionPrescribing patterns for paediatric hyperopia vary significantly among eye care providers.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.