At the ultra-thin limit below 20 nm, a film's electrical conductivity, thermal conductivity, or thermoelectricity depends heavily on its thickness. In most studies, each sample is fabricated one at a time, potentially leading to considerable uncertainty in later characterizations. We design and build an in-situ apparatus to measure thermoelectricity during their deposition inside a thermal evaporator. A temperature difference of up to 2 K is generated by a current passing through an on-chip resistor patterned using photolithography. The Seebeck voltage is measured on a Hall bar structure of a film deposited through a shadow mask. The measurement system is calibrated carefully before loading into the thermal evaporator. This in-situ thermoelectricity measurement system has been thoroughly tested on various materials, including Bi, Te, and Bi2Te3, at high temperatures up to 500 K.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.