Objectives
Novel automated computed tomography (CT) software (Valve ASSIST 2) has been developed for transcatheter aortic valve implantation (TAVI), which not only provides three‐dimensional (3D) reconstruction of multidetector (MD) CT images, but also enables intraprocedural real‐time fusion of fluoroscopic and MDCT images. We aimed to clarify the reproducibility and accuracy of this software in the aortic annulus assessment and verify the potential of intraprocedural integrated MDCT imaging for TAVI.
Methods and results
We examined 50 patients with severe aortic stenosis undergoing transfemoral TAVI. Aortic annulus measurements were performed using 3mensio and the novel planning software. For intraprocedural imaging, preoperative CT dataset was overlaid onto fluoroscopy with the fusion software. The two images were aligned using the aortic root anatomy visible on both modalities. Novel planning software provided excellent reproducibility for the measurement of aortic annulus area (intraobserver intraclass correlation coefficients [ICC] 0.959, interobserver ICC 0.941), and perimeter (intraobserver ICC 0.915, interobserver ICC 0.912). Excellent correlation was found between novel planning software and 3mensio (ICC 0.952 for aortic annulus area, and 0.923 for perimeter). Intraprocedural fusion image of CT aortography and fluoroscopic aortic root aortography generated by this novel software identified coronary orifices and the distribution of aortic valve calcification during the device positioning. Fusion image displayed coronary orifices after device implantation.
Conclusions
Novel planning software showed excellent reproducibility and accuracy in the assessment of aortic root anatomy. Furthermore, the integrated 3D fusion image might have a potential as an intraprocedural imaging modality to contribute to the development of a safer TAVI procedure.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.