The fabrication of toner-based microfluidic devices to perform clinical diagnostics with capillary action and colorimetric detection is described in this report. Test zones and microfluidic channels were drawn in a graphic software package and laser printed on a polyester film. The printed layout and its mirror image were aligned with an intermediary cut-through polyester film and then thermally laminated together at 150 °C at 60 cm/min to obtain a channel with ca. 100-μm depth. Colorimetric assays for glucose, protein, and cholesterol were successfully performed using a desktop scanner. The limit of detection (LD) values found for protein, cholesterol, and glucose were 8, 0.2, and 0.3 mg/mL, respectively. The relative standard deviation (RSD) values for an interdevices comparison were 6%, 1%, and 3% for protein, cholesterol, and glucose, respectively. Bioassays were successfully performed on toner-based devices stored at different temperatures during five consecutive days without loss of activity.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.