Objectives: To evaluate an implanted thermal ablation device that can be heated with high efficiency using a resonant circuit as the implant. Methods: 16 rats were used. The implants, adjusted at a resonance frequency of 4 MHz, were fixed on the surface of the liver of rats under laparotomy. In 14 of 16 rats, an alternating magnetic field (AMF) was applied for 6 min with an output of 300 W from outside the body using a ferrite core applicator. The implant temperature during AMF exposure was measured. The 14 rats were divided into 5 groups, depending on time from AMF application until they were sacrificed (1 h, 1 day, 3 days, 7 days and 1 month after application). Two rats not exposed to AMF were used as controls. Livers were removed and evaluated; the cross-sectional area and width of the ablated region were measured. Results: During AMF exposure, the implant temperature rose to 127.8¡39.3 u C (mean ¡ standard deviation). The cross-sectional area of the ablated region was largest after 1 day and tended to decrease with time. The widths of the ablated region were 4.87¡0.22 mm, 4.15¡0.36 mm, 3.67¡0.58 mm and 3.24¡0.16 mm in the 1 day, 3 day, 7 day and 1 month groups, respectively. No significant differences (p,0.05) were seen in either cross-sectional area or width of the ablated region. Conclusion: Sufficient heat for ablation was obtained in vivo using a newly developed implanted thermal ablation device. This device may be a new option for thermal ablation therapy.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.