Novel electrochemical methods have been developed for determination of total hemoglobin, hematocrit, and detection of hemolysis in whole blood. Hemoglobin is measured through its peroxidase activity, a fluoride ion-selective electrode being used to monitor the rate of fluoride ion production from the oxidation of an organofluorine compound. Results agree well with those obtained with the cyanmethemoglobin method (r = 0.970). Hematocrit is determined from the ratio of the sodium ion concentrations measured with an ion-selective electrode before and after lysis of the erythrocytes. Results by this and the microhematocrit method correlated well (r = 0.987). Hemolysis in a whole-blood sample is detected by using an oxygen electrode to measure the oxygen released when hemoglobin in plasma is oxidized.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.