Basal ganglia involvement is a typical finding in patients with neurological complications of HUS. Even in patients with severe CNS involvement on acute imaging studies, prognosis was favourable for clinical outcome and resolution of pathological imaging findings.
Normal values for ONSD in children and adolescents found in this study are significantly higher than assumed. The values found for transbulbar sonography are confirmed by comparable results for MR measurements. A precise sonographic measurement technique and the consideration of normal values found hereby are essential for correct interpretation of ONSD measurements in children and adolescents.
PURPOSE: To evaluate the accuracy of measurements of the optic nerve sheath diameter (ONSD) in children by comparing transbulbar sonography with magnetic resonance imaging (MRI). MATERIALS AND METHODS: 65 children (age: 3 mo-17 y; mean age: 11.3 y) underwent MR imaging of the brain including a heavily T 2-w sequence of the orbit and transbulbar sonography using a 17 MHz linear array transducer. Measurements of the ONSD were performed retrospectively by two experienced readers and all images were evaluated regarding the image quality. Bland-Altman Plots were produced to assess the accuracy of measurements. The correlation between readers and between MR imaging and transbulbar sonography was calculated by the concordance correlation coefficient (CCC). RESULTS: Overall the mean values of the ONSD for MRI (5.86 ± 0.66 mm) and transbulbar sonography (5.86 ± 0.71 mm) were identical. There was a high correlation between readers for measurements of the ONSD both for transbulbar sonography (CCC = 0.93) as well as for MRI (CCC = 0.9). Comparing the measurement values between transbulbar sonography and MRI, the correlation of ONSD values was good to moderate with a high dependency on image quality (CCC [0.31, 0.68]). CONCLUSION: We were able to demonstrate that the ONSD values of transbulbar sonography in children correlate well with MRI if the relevant anatomic structures are depicted and the measuring points are set correctly. Based on the findings of our study, it seems necessary to define normal and cut-off values for the ONSD in children again to finally assess the clinical relevance of the ONSD as a noninvasive parameter for the evaluation of intracranial pressure.
If decisive anatomic structures are clearly depicted and the measuring points are set correctly, there is a good correlation between HRUS and MRI measurements of the optic nerve and the ONSD even on transbulbar sonography. As most of the standard and cut-off values that have been published for ultrasound are significantly lower than the results obtained with MRI, a reevaluation of sonographic ONSD measurement with correlation to MRI is necessary.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.