Purpose
The purpose of this study is to provide a standard method for flow velocity measurements with phase‐contrast (PC) MRI. This method can be used for in vitro studies that place high demands on measurement accuracy. Clinically relevant PC MRI techniques can be validated using this method before being applied in vivo.
Methods
Many motion‐related errors in PC MRI, particularly flow misregistration, depend on the timing of the encoding gradients in the pulse sequence. By synchronizing all encoding gradients and shortening the overall encoding interval, these errors can be significantly reduced. Based on this concept, a single‐point PC MRI method is proposed.
Results
Flow experiments were conducted in vitro. No considerable errors were found in the velocity data of the proposed method. For comparison, a conventional PC MRI technique showed up to 100% local velocity deviation and up to 35% flow rate deviation in the same experiments.
Conclusions
With the proposed method, the overall measurement accuracy is significantly increased compared to conventional PC MRI techniques. Due to long acquisition times and high specific absorption rates, this method can only be applied in vitro.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.