Objective A clinical investigation into a new bone void filler is giving
first data on systemic and local exposure to the anti-infective
substance after implantation.Method A total of 20 patients with post-traumatic/post-operative bone
infections were enrolled in this open-label, prospective study.
After radical surgical debridement, the bone cavity was filled with
this material. The 21-day hospitalisation phase included determination
of gentamicin concentrations in plasma, urine and wound exudate, assessment
of wound healing, infection parameters, implant resorption, laboratory
parameters, and adverse event monitoring. The follow-up period was
six months. Results Systemic exposure to gentamicin after implantation was very low
as local gentamicin concentrations were measured in wound exudate
after six to ten hours. There were no signs of infectious complication
throughout the clinical phase. Four patients had recurrent infections
several weeks to months after implantation. The outcome was deemed successful
by remission of infection in 16 (80%) of these problematic long-term
treated patients. Safety laboratory measurements did not indicate
nephrotoxic or hepatotoxic effects. Conclusions Local application of calcium sulphate/carbonate bone void filler
comprising gentamicin revealed sufficient active local levels of
the antibiotic by simultaneous significant low systemic exposure
in patients with mostly chronic osteomyelitis/osteitis. The material
was safe and well tolerated.Cite this article: Bone Joint Res 2014;3:223–9.
IntroductionIn revision total hip arthroplasty (THA), the cancellous bone is normally completely removed out of the femoral canal during stem extraction. This situation is comparable to primary THA following the shape-closed concept, with some authors advocating to remove the metaphyseal cancellous bone to enhance press-fit stability (“French paradox”). The aim of this study was to investigate the long-term outcome, regarding survival and radiological results, of a cemented straight stem when used for revision THA and to compare these results to the results of the same stem in primary THA.Materials and methods178 stem revisions performed between 01/1994 and 08/2008 using the Virtec straight stem were included. The cumulative incidence for re-revision was calculated using a competing risk model. Risk factors for re-revision of the stem were analyzed using an absolute risk regression model. Radiographs analyzed for osteolysis, debonding and subsidence had a minimum follow-up of 10 years.ResultsThe cumulative incidence for re-revision due to aseptic loosening of the stem was 5.5% (95% CI, 2.9–10.2%) at 10 years. Aseptic loosening was associated with younger age, larger defect size and larger stem size. After a minimum 10-year follow-up, osteolysis was seen in 39 of 80 revision THA. Compared to the results in primary THA, the survival in revision THA with the same implant was inferior.ConclusionsCemented straight stems used for revision THA showed excellent long-term results regarding survivorship and radiological outcome. This stem therefore offers a valuable and cost-effective option in revision THA.
Persistent osteomyelitis is a severe and challenging problem in bone surgery. We describe a surgical intervention in a young adult which combines a bone debridement process, a plastic muscle flap sealing and the administration of a novel bone substitution material with anti-infective properties. After 1 year, the patient showed no signs or symptoms of a reoccurrence of infection with full load capacity of the treated leg.
Slow absorption of both antibiotics after release from the cement resulted in plasma concentrations well below toxic levels and did not result in a critical systemic concentration potientially inducing bacterial resistance. The treatment with this novel bone cement was assessed as efficacious and was very well tolerated by all patients.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.