We evaluated a new protocol for measurement of cyclosporine A (CsA) 2 H after dose (C2) on the V-Twin® analyzer. Imprecision, recovery, and linearity were determined using CsA-spiked blood pools. Accuracy was evaluated using specimens from renal, cardiac, and liver transplant patients, and results were compared with those from liquid chromatography-tandem mass spectrometry (LC-MS/MS) and the Abbott TDx®/TDxFLx® assay. Cross-reactivity and interferences were assessed in the presence of 800 ng/mL CsA. Imprecision coefficients of variation were 3.3%-4.8% (within run) and 5.9%-8.7% (total). Recovery was within 10% of the expected values. Linearity was 350-2,000 ng/mL. Calibration was stable for ≥ 2 weeks. Method comparison showed regression statistics: V-Twin® = 1.01 × LC tandem MS + 36.1, r = 0.971; V-Twin® = 1.13 × Abbott - 92.4, r = 0.969. Metabolite cross-reactivity and interference (endogenous substances and drugs) were within ±10%. The C2 protocol on the V-Twin® analyzer provides acceptable assay performance and accurate determination of whole blood CsA drawn at 2 H after dose.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.