Disregarding local coils in PET AC can lead to a bias of the AC PET images that is regional dependent. The closer the analyzed region is located to the coil, the higher the bias. Cod liver oil capsules or the UTE sequence can be used for RF coil position determination. The middle part of the examined RF coil hosting the preamplifiers and electronic components provides the highest attenuating part. Consequently, emphasis should be put on correcting for this portion of the RF coils with the suggested methods.
The successful integration of a four-channel RF breast coil with a defined table position together with the CT-based μ-maps provides a technical basis for future clinical PET/MR breast imaging applications.
The presence of flexible RF surface coils leads to considerable local errors in the simultaneously measured PET activity concentration up to 15.5% especially in regions close to the coils. The presented automatic algorithm accurately and reliably reduces the PET quantification errors caused by multiple partly overlapping flexible RF surface coils to values of 4.3% or better.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.