Background-A shortage of donor organs and increased numbers of deaths of patients on the waiting list for cardiac transplantation make mechanical circulatory support for a bridge to transplantation a standard clinical procedure. Continuous-flow rotary blood pumps offer exciting new perspectives. Methods and Results-Two male patients (ages 44 and 65 years) suffering from end-stage left heart failure were implanted with a DeBakey VAD axial-flow pump for use as a bridge to transplant. In the initial postoperative period, the mean pump flow was 3.9Ϯ0.5 L/min, which equals a mean cardiac index (CI) of 2.3Ϯ0.2 L ⅐ min Ϫ1 ⅐ m
An accurate characterization of the hemodynamic behavior of ventricular assist devices (VADs) is of paramount importance for proper modeling of the heart-pump interaction and the validation of control strategies. This paper describes an advanced test bench, which is able to generate complex hydraulic loads, and a procedure to characterize rotary blood pump performance in a pulsatile environment. Special focus was laid on model parameter identifiability in the frequency domain and the correlation between dynamic and steady-state models.
Twelve combinations of different flow/head/speed signals, which covered the clinical VAD working conditions, were generated for the pump characterization. Root mean square error (RMSE) between predicted and measured flow was used to evaluate the VAD model. The found parameters were then validated with broadband random signals.
In the experiments the optimization process always successfully converged. Even in the most demanding dynamic conditions the RMSE was 7.4 ml/sec and the absolute error never exceeded 24.9 ml/sec. Validity ranges for the identified VAD model were: flow 0–180 ml/sec; head 0–120 mmHg; speed 7.5–12.5 krpm.
In conclusion, a universal test bench and a characterization procedure to describe the hydrodynamic properties of rotary blood pumps were established. For a particular pump, a reliable mathematical model was identified that correctly reproduced the relationship between instantaneous VAD flow, head and impeller speed.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.