Our data enlarge the body of evidence showing that ROTEM(®) assays are useful in trauma patients. Treatment concepts should focus on maintaining fibrin polymerization and treating hyperfibrinolysis.
Loss of HMW vWF multimer bands occurred in patients undergoing ECMO support and resolved after the termination of ECMO. Although not detectable with coagulation screening tests, a vWF:RCo/vWF:Ag ratio <0.7 during ECMO was highly indicative for loss of HMW vWF multimers. Our findings may at least in part explain increased bleeding tendency during ECMO therapy. Administration of vWF concentrates may support restoration of primary hemostasis in patients with relevant bleeding during ECMO support.
The results of this study demonstrated that increased creatinine levels and lower PT before ECMO and secondary impaired platelet function significantly increased transfusion requirement.
Thromboelastometry point-of-care coagulation testing facilitates optimised management of bleeding. Previous thromboelastometry systems required the blood sample and liquid reagents to be pipetted in several manual steps by trained personnel. The ROTEMsigma coagulation analyser is a fully automated point-of-care device. We aimed to assess the reference ranges of the new device and to compare the results with those of the predecessor device, the ROTEMdelta. We took blood from healthy volunteers and from hyper-or hypocoagulable patients; blood samples from healthy volunteers served to determine reference ranges for the most important parameters for the ROTEMsigma: CT EXTEM 48-61 s; A5 EXTEM 30-51 mm; MCF EXTEM 54-70 mm; CT INTEM 138-174 s; MCF INTEM 51-67 mm and MCF FIBTEM 5-24 mm. We then used blood samples from patients to compare the results obtained between the old and the new device. We found a strong correlation between the same tests performed on two ROTEMsigma devices and between the ROTEMsigma and the ROTEMdelta with respect to the determination of thromboelastometry parameters of hyper-and hypocoagulable patients (all p < 0.001 and R > 0.8). Performance evaluation for the ROTEMsigma device showed very high precision (R > 0.99, p < 0.001). Our reference ranges can serve as an important aid for other hospitals using this new device.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.