Bi-modular hip arthroplasty prostheses allow adaptation to the individual patient anatomy and the combination of different materials but introduce an additional interface, which was related lately to current clinical issues. Relative motion at the additional taper interface might increase the overall risk of fretting, corrosion, metallic debris and early failure. The aim of this study was to investigate whether the assembly force influences the relative motion and seating behaviour at the stem-neck interface of a bi-modular hip prosthesis (Metha(®); Aesculap AG, Tuttlingen, Germany) and whether this relation is influenced by the taper angle difference between male and female taper angles. Neck adapters made of titanium (Ti6Al4V) and CoCr (CoCr29Mo) were assembled with a titanium stem using varying assembly forces and mechanically loaded. A contactless eddy current measurement system was used to record the relative motion between prosthesis stem and neck adapter. Higher relative motion was observed for Ti neck adapters compared to the CoCr ones (p < 0.001). Higher assembly forces caused increased seating distances (p < 0.001) and led to significantly reduced relative motion (p = 0.019). Independent of neck material type, prostheses with larger taper angle difference between male and female taper angles exhibited decreased relative motion (p < 0.001). Surgeons should carefully use assembly forces above 4 kN to decrease the amount of relative motion within the taper interface. Maximum assembly forces, however, should be limited to prevent periprosthetic fractures. Manufacturers should optimize taper angle differences to increase the resistance against relative motion.
The stem-neck taper interface of bimodular hip endoprostheses bears the risk of micromotions that can result in ongoing corrosion due to removal of the passive layer and ultimately cause implant fracture. We investigated the extent of micromotions at the stem-neck interface and the seating behavior of necks of one design made from different alloys during daily activities. Modular hip prostheses (n ¼ 36, Metha1, Aesculap AG, Germany) with neck adapters (CoCr29Mo6 or Ti6Al4V) were embedded in PMMA (ISO 7206-4) and exposed to cyclic loading with peak loads ranging from walking (F max ¼ 2.3 kN) to stumbling (F max ¼ 5.3 kN). Translational and rotational micromotions at the taper interface and seating characteristics during assembly and loading were determined using four eddy-current sensors. Seating during loading after implant assembly was dependent on load magnitude but not on material coupling. Micromotions in the stem-neck interface correlated positively with load levels (CoCr: 2.6-6.3 mm, Ti: 4.6-13.8 mm; p < 0.001) with Ti neck adapters exhibiting significantly larger micromotions than CoCr (p < 0.001). These findings explain why high body weights and activities related to higher loads could increase the risk of fretting-induced implant failures in clinical application, especially for Ti-Ti combinations. Still, the role of taper seating is not clearly understood. ß
Fretting corrosion is one contributor to the clinical failure of modular joint arthroplasty. It is initiated by micromotion in metal junctions exposed to fluids. Omitting metal-on-metal contacts could help to reduce the corrosion risk. The coating of one metal taper partner with a ceramic-based silicon nitride (SiNx) coating might provide this separation. The aim of the study was to identify whether a SiNx coating of the male taper component influences the micromotion within a taper junction. Hip prosthesis heads made of CoCr29Mo6 (Aesculap) and Ti6Al4V (Peter Brehm) were assembled (2000 N) to SiNx-coated and uncoated stem tapers made of Ti6Al4V and CoCr29Mo6 (2×2×2 combinations, each n = 4). Consecutive sinusoidal loading representing three daily activities was applied. Contactless relative motion in six degrees of freedom was measured using six eddy-current sensors. Micromotion in the junction was determined by compensating for the elastic deformation derived from additional monoblock measurements. After pull-off, the taper surfaces were microscopically inspected. Micromotion magnitude reached up to 8.4 ± 0.8 µm during loading that represented stumbling. Ti6Al4V stems showed significantly higher micromotion than those made of CoCr29Mo6, while taper coating had no influence. Statistical differences in pull-off forces were found for none of the taper junctions. Microscopy revealed CoCr29Mo6 abrasion from the head taper surface if combined with coated stem tapers. Higher micromotion of Ti6Al4V tapers was probably caused by the lower Young’s modulus. Even in the contact areas, the coating was not damaged during loading. The mechanics of coated tapers was similar to uncoated prostheses. Thus, the separation of the two metal surfaces with the objective to reduce in vivo corrosion appears to be achievable if the coating is able to withstand in vivo conditions. However, the hard ceramic-based stem coating lead to undesirable debris from the CoCr29Mo6 heads during loading.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.