A thermoluminescence dosemeter (TLD)-entrance surface dose (ESD) phantom combination was calibrated in terms of air-kerma in IEC RQR X-ray radiation qualities between 50 and 150 kV. The ESD phantom was designed and constructed as a part of the work. With the combination, air-kermas were measured for four radiological examinations (two chest and two lumbar spine examinations in two hospitals), and converted to ESDa using Monte-Carlo calculated BSF data tabulated for different tube voltages, filtrations and beam diameters at 1 m distance. The results agreed with the ionisation-chamber measurements within the reported overall uncertainty of the TLD method. In the calibration, the ESD phantom can be replaced by the ISO water slab phantom unaffecting the reported overall uncertainty. Backscatter-related parameters for the ISO water slab phantom and the newly designed ESD phantom were determined for the IEC RQR qualities used in the secondary standard dosimetry laboratory of STUK at 1 m distance, including an approximate Hp(10)/Hp(0.07) ratio.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.