A suppressed ion method for the determination method of eight organic acids (formic, malonic, lactic, acetic, malefic, citric, fumaric, and propionic acid) has been studied for reversedphase high-performance liquid chromatography (RP-HPLC) with a new triacontyl groups (C30) column, coupled with an ultraviolet detector (UV) at 210 nm. A 50 mM ammonium dihydrogenphosphate (NH4H2P04) aqueous solution (pH 2.0) containing phosphoric acid (H3PO4) was used as an optimal mobile phase for the separation of eight organic acids. A good linearity of the calibration curve was obtained in the concentration range of 5 mgl-1 to 100 mgl-1(1|r| =0.9999-0.9994). The detection limit based on S/N=3 were 3.8 mgl-1 for formic acid, 3.5 mgl-1 for malonic acid, 3.1 mg l-1 for lactic acid, 3.4 mgl-1 for acetic acid, 1.1 mgl-1 for maleic acid, 3.4 mgl-1 for citric acid, 3.3 mgl-1 for fumaric acid, and 3.6 mgl-1 for propionic acid. This analytical method was successfully applied to the determination of eight organic acids in sake and wine.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.