Purpose: In clinical settings, ultrasonography (US) has recently been used to aid in the insertion of peripheral intravenous catheters (PIVCs). This cross-sectional study aimed to verify the reliability and validity of a tablet-type device in assessing vein size and depth for catheter site selection and detecting thrombus with resultant subcutaneous edema as a cause of catheter failure using US.
Methods: Adult patients receiving infusions via a PIVC at a university hospital between January and February 2017 were included. All participants underwent US at the PIVC site. An expert sonographer and a nurse blinded to all information, except for ultrasonograms, evaluated the data. Intraclass correlation with 95% confidence interval (CI) was used to evaluate interrater and intrarater reliability of US assessment. To assess criterion-related validity, a high-end US notebook device was used for reference data collection. Pearson's correlation coefficient was used to evaluate criterion-related validity.
Results: We observed 21 patients with 26 catheters. Intraclass correlations (95% CI) for the measured vein diameters and depths were as follows: intrarater reliability, 0.92 (0.57–0.98) and 0.78 (0.10–0.95); interrater reliability, 0.95 (0.78–0.99) and 0.94 (0.77–0.99); and Pearson's correlation coefficient for criterion-related validity, 0.74 (P = 0.02) and 0.77 (P = 0.02), respectively. However, the analysis of causes of catheter failure did not show reliable validity.
Conclusion: This pilot study suggests that the tablet-type device is useful for assessing peripheral veins in clinical settings.
The frequency of phlebitis was lower in the polyurethane, in which the catheter was placed at lower angle, almost parallel to the vessel. Our results will aid in developing new catheters and in improving PIVC-securement techniques.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.