The study aimed to compare the dose distribution in permanent low-dose-rate brachytherapy (LDR-BT) and high-dose-rate brachytherapy (HDR-BT), specifically focusing on the impact of a spacer and prostate volume. The relative dose distribution of 102 LDR-BT patients (prescription dose 145 Gy) at different intervals was compared with the dose distribution of 105 HDR-BT patients (232 HDR-BT fractions with prescription doses of 9 Gy, n = 151, or 11.5 Gy, n = 81). A hydrogel spacer (10 mL) was only injected before HDR-BT. For the analysis of dose coverage outside the prostate, a 5 mm margin was added to the prostate volume (PV+). Prostate V100 and D90 of HDR-BT and LDR-BT at different intervals were comparable. HDR-BT was characterized by a considerably more homogenous dose distribution and lower doses to the urethra. The minimum dose in 90% of PV+ was higher for larger prostates. As a consequence of the hydrogel spacer in HDR-BT patients, the intraoperative dose at the rectum was considerably lower, especially in smaller prostates. However, prostate volume dose coverage was not improved. The dosimetric results well explain clinical differences between these techniques reported in the literature review, specifically comparable tumor control, higher acute urinary toxicity rates in LDR-BT in comparison to HDR-BT, decreased rectal toxicity after spacer placement, and improved tumor control after HDR-BT in larger prostate volumes.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.