A novel generic approach based on precolumn isotope dilution nanoHPLC-ICPMS analysis was developed for the accurate absolute quantification of sulfur-containing peptides. A 34S-labeled, species-unspecific sulfur spike (sulfate), noninteracting with analyte peptides under the optimized HPLC condition, was added directly to the chromatographic eluents. Thus a generic sulfur standard permanently present during analysis was used for peptide quantification. Interference-free detection of the 32S and 34S isotopes in ICPMS was achieved by eliminating O2+ ions in a collision cell using Xe gas at 130 microL min-1. The detection limit for sulfur was 45 microg L-1 which corresponded to 1-2 pmol of individual peptides. The method was validated by the analysis of a standard peptide solution showing high accuracy (recovery 103%) and good precision (RSD 2.1%). The combination of nanoHPLC-ICP IDMS with nanoHPLC-ESI MS/MS allowed the precise quantification and identification of sulfur-containing peptides in tryptic digests of human serum albumin and salt-induced yeast protein (SIP18) at the picomole level.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.