This paper investigates different possibilities of measurement and control of pressure in the cryostat filled with liquid nitrogen. Laboratory designed membrane pressure regulator for automatic control of pressure in the range from 0.9 to 1.3 bar gauge in the cryostat is presented. The performance of the special purpose regulator is evaluated through measurement of pressure deviations from set-point value and their effect on thermal conditions in the cryostat. Results are compared to values obtained using manual pressure regulator.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.