Background: The population of children with chronic respiratory failure requiring long-term mechanical ventilation in the home has grown worldwide. The optimal choice from an increasing number of home ventilators commercialized for children is often challenging for the attending physicians. The aim of the present study was to compare the trigger performance of five pediatric bilevel home ventilators and one intensive care unit ventilator depending on circuit type and system leak.
Methods:The trigger performances of the ventilators were compared in combination with all compatible circuits using a physical model of the lung with increasing system leak. The flow generator simulated the patient's breathing effort with flow rates of 2.7-6.4 L/min at a frequency of 30 breaths/min. All ventilators were set to deliver 16 cmH 2 O inspiratory pressure support and 4 cmH 2 O positive endexpiratory pressure.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.