IntroductionRadiation dose to patients undergoing invasive coronary angiography (ICA) is relatively high. Guidelines suggest that a local benchmark or diagnostic reference level (DRL) be established for these procedures. This study sought to create a DRL for ICA procedures in Queensland public hospitals.MethodsData were collected for all Cardiac Catheter Laboratories in Queensland public hospitals. Data were collected for diagnostic coronary angiography (CA) and single-vessel percutaneous intervention (PCI) procedures. Dose area product (PKA), skin surface entrance dose (KAR), fluoroscopy time (FT), and patient height and weight were collected for 3 months. The DRL was set from the 75th percentile of the PKA.Results2590 patients were included in the CA group where the median FT was 3.5 min (inter-quartile range = 2.3–6.1). Median KAR = 581 mGy (374–876). Median PKA = 3908 uGym2 (2489–5865) DRL = 5865 uGym2. 947 patients were included in the PCI group where median FT was 11.2 min (7.7–17.4). Median KAR = 1501 mGy (928–2224). Median PKA = 8736 uGym2 (5449–12,900) DRL = 12,900 uGym2.ConclusionThis study established a benchmark for radiation dose for diagnostic and interventional coronary angiography in Queensland public facilities.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.