The auxiliary calibration device has shown calibration results statistically equal to those obtained with a reference device in the laboratory environment. We have found less variability of hemolysis rates in the operating room, using the auxiliary device and we did not observe any influence of unbalanced rollers in the hemolysis rates.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.