Metallic implants used in bone and joint arthroplasty induce severe spatial perturbations to the B 0 magnetic field used for high-field clinical magnetic resonance. These perturbations distort slice-selection and frequency encoding processes applied in conventional two-dimensional MRI techniques and hinder the diagnosis of complications from arthroplasty. Here, a method is presented whereby multiple three-dimensional fast-spin-echo images are collected using discrete offsets in RF transmission and reception frequency. It is demonstrated that this multi acquisition variable-resonance image combination technique can be used to generate a composite image that is devoid of slice-plane distortion and possesses greatly reduced distortions in the readout direction, even in the immediate vicinity of metallic implants.Magn Reson Med 61:381-390, 2009.
The proposed phantom and protocol can be an effective tool for verifying the spatial accuracy of an MR system, which in turn can improve the accuracy and confidence of MR guided therapies. Data from this protocol may also be used in the development of advanced distortion correction algorithms. Employed by General Electric Healthcare.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.