The current and potential applications of atmospheric pressure plasmas in medicine generate an increasing need to develop safe and reliable plasma devices for patient treatment. This paper shows how the estimation of safety risks, the stability of the generated plasma, and the effectiveness in the aimed application can orientate the design process of a specific atmospheric pressure plasma device intended for clinical use. A promising plasma jet device operated with air is optimized, leading to a configuration with a more advanced design that reduces the temperature of the effluent, prevents the material degradation and improves the isolation of the high voltage components. The effects of the plasma jet treatment are investigated by chemical analysis of demineralized water and inactivation tests on E. coli cultures.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.