Trapezium components from two uncemented total joint replacements were compared in a three-dimensional finite element model. A 100 N axial and angular load was applied in a normal and an osteoporotic bone model. The axial deformation and maximum periprosthetic stress are greater for the Elektra than the Motec CMC cup. The Motec CMC design is less sensitive to changing bone quality. The Elektra cup transmits more stress to the cortical bone rim in all load conditions, but under angular loading the proportionate increase in stress is lower. The Motec CMC design distributes the stress and contact pressure more evenly, whereas the Elektra transfers most of the load to the cortical bone rim and the screw hole base. The design features that are believed to be of greatest significance for the differences are the raised centre of rotation of the Motec CMC cup and the collar acting as a lever arm.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.