We describe use of a bidirectional neuromyoelectric prosthetic hand that conveys biomimetic sensory feedback. Electromyographic recordings from residual arm muscles were decoded to provide independent and proportional control of a six-DOF prosthetic hand and wrist—the DEKA LUKE arm. Activation of contact sensors on the prosthesis resulted in intraneural microstimulation of residual sensory nerve fibers through chronically implanted Utah Slanted Electrode Arrays, thereby evoking tactile percepts on the phantom hand. With sensory feedback enabled, the participant exhibited greater precision in grip force and was better able to handle fragile objects. With active exploration, the participant was also able to distinguish between small and large objects and between soft and hard ones. When the sensory feedback was biomimetic—designed to mimic natural sensory signals—the participant was able to identify the objects significantly faster than with the use of traditional encoding algorithms that depended on only the present stimulus intensity. Thus, artificial touch can be sculpted by patterning the sensory feedback, and biologically inspired patterns elicit more interpretable and useful percepts.
We quantified prosthesis embodiment and phantom pain reduction associated with motor control and sensory feedback from a prosthetic hand in one human with a long-term transradial amputation. Microelectrode arrays were implanted in the residual median and ulnar arm nerves and intramuscular electromyography recording leads were implanted in residual limb muscles to enable sensory feedback and motor control. Objective measures (proprioceptive drift) and subjective measures (survey answers) were used to assess prosthesis embodiment. For both measures, there was a significant level of embodiment of the physical prosthetic limb after open-loop motor control of the prosthesis (i.e., without sensory feedback), open-loop sensation from the prosthesis (i.e., without motor control), and closed-loop control of the prosthesis (i.e., motor control with sensory feedback). There was also a statistically significant reduction in reported phantom pain after experimental sessions that included open-loop nerve microstimulation, open-loop prosthesis motor control, or closed-loop prosthesis motor control. The closed-loop condition provided no additional significant improvements in phantom pain reduction or prosthesis embodiment relative to the open-loop sensory condition or the open-loop motor condition. This study represents the first long-term (14-month), systematic report of phantom pain reduction and prosthesis embodiment in a human amputee across a variety of prosthesis use cases.
Background: Multi-articulate prostheses are capable of performing dexterous hand movements. However, clinically available control strategies fail to provide users with intuitive, independent and proportional control over multiple degrees of freedom (DOFs) in real-time. New Method: We detail the use of a modified Kalman filter (MKF) to provide intuitive, independent and proportional control over six-DOF prostheses such as the DEKA "LUKE" Arm. Input features include neural firing rates recorded from Utah Slanted Electrode Arrays and mean absolute value of intramuscular electromyographic (EMG) recordings. Ad-hoc modifications include thresholds and nonunity gains on the output of a Kalman filter. Results: We demonstrate that both neural and EMG data can be combined effectively. We also highlight that modifications can be optimized to significantly improve performance relative to an unmodified Kalman filter. Thresholds significantly reduced unintended movement and promoted more independent control of the different DOFs. Gain were significantly greater than one and served to amplify participant effort. Optimal modifications can be determined quickly offline and translate to functional improvements online. Using a portable take-home system, participants performed various activities of daily living. Comparison with Existing Methods: In contrast to pattern recognition, the MKF allows users to continuously modulate their force output, which is critical for fine dexterity. The MKF is also computationally efficient and can be trained in less than five minutes. Conclusions: The MKF can be used to explore the functional and psychological benefits associated with long-term, at-home control of dexterous prosthetic hands. HIGHLIGHTS:• Neural and electromyographic signals can be used together to estimate motor intent • Ad-hoc thresholds and gains improve prosthetic control by reducing unintended movement • Thresholds and gains can be optimized offline and translate to functional improvements online
Objective. We explore the long-term performance and stability of seven percutaneous Utah Slanted Electrode Arrays (USEAs) and intramuscular recording leads (iEMGs) implanted chronically in the residual arm nerves and muscles of three human participants as a means to permanently restore sensorimotor function after transradial amputations. Approach. We quantify the number of functional recording and functional stimulating electrodes over time. We also calculate the signal-to-noise ratio (SNR) of USEA and iEMG recordings and quantify the stimulation current necessary to evoke detectable sensory percepts. Furthermore, we quantify the consistency of the sensory modality, receptive field location, and receptive field size of USEA-evoked percepts. Main results. In the most recent subject, involving USEAs with technical improvements, neural recordings persisted for 502 d (entire implant duration) and the number of functional recording electrodes for one USEA increased over time. However, for six out of seven USEAs across the three participants, the number of functional recording electrodes decreased within the first 2 months after implantation. The SNR of neural recordings and electromyographic recordings stayed relatively consistent over time. Sensory percepts were consistently evoked over the span of 14 months, were not significantly different in size, and highlighted the nerves’ fascicular organization. The percentage of percepts with consistent modality or consistent receptive field location between sessions (∼1 month apart) varied between 0%–86.2% and 9.1%–100%, respectively. Stimulation thresholds and electrode impedances increased initially but then remained relatively stable over time. Significance. This work demonstrates improved performance of USEAs, and provides a basis for comparing the longevity and stability of USEAs to that of other neural interfaces. USEAs provide a rich repertoire of neural recordings and sensory percepts. Although their performance still generally declines over time, functionality can persist long-term. Future work should leverage the results presented here to further improve USEA design or to develop adaptive algorithms that can maintain a high level of performance.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.
customersupport@researchsolutions.com
10624 S. Eastern Ave., Ste. A-614
Henderson, NV 89052, USA
This site is protected by reCAPTCHA and the Google Privacy Policy and Terms of Service apply.
Copyright © 2024 scite LLC. All rights reserved.
Made with 💙 for researchers
Part of the Research Solutions Family.