Nickel(II) and lead(II) ionic imprinted 8-hydroxyquinoline polymers were synthesized by a precipitation polymerization technique and were used as selective solid phase extraction supports for the determination of nickel and lead in seawater by flow injection solid phase extraction on-line inductively coupled plasma-optical emission spectrometry. An optimum loading flow rate of 2.25 mL min(-1) for 2 min and an elution flow rate of 2.25 mL min(-1) for 1 min gave an enrichment factor of 15 for nickel. However, a low dynamic capacity and/or rate for adsorption and desorption was found for lead ionic imprinted polymer and a flow rate of 3.00 mL min(-1) for 4-min loading and a flow rate of 2.25 mL min(-1) for 1-min elution gave a enrichment factor of 5. The limit of detection was 0.33 microg L(-1) for nickel and 1.88 microg L(-1) for lead, with a precision (n = 11) of 8% (2.37 microg Ni L(-1)) for nickel and 11% (8.38 microg Pb L(-1)) for lead. Accuracy was also assessed by analyzing SLEW-3 (estuarine water) and TM-24 (lake water) certified reference materials, and the values determined were in good agreement with the certified concentrations.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.