This paper proposes a manipulability optimization control of a 7-DoF robot manipulator for Robot-Assisted Minimally Invasive Surgery (RAMIS), which at the same time guarantees a Remote Center of Motion (RCM). The first degree of redundancy of the manipulator is used to achieve an RCM constraint, the second one is adopted for manipulability optimization. A hierarchical operational space formulation is introduced to integrate all the control components, including a Cartesian compliance control involving the main surgical task, a first null-space controller for the RCM constraint, and a second null-space controller for manipulability optimization. Experiments with virtual surgical tasks, in an augmented reality environment, were performed to validate the proposed control strategy using the KUKA LWR4+. The results demonstrate that end-effector accuracy and RCM constraint can be guaranteed, along with improving the manipulability of the surgical tip.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.