BackgroundPatients with deep brain stimulation (DBS) implants have limited access to MRI due to safety concerns associated with RF‐induced heating. Currently, MRI in these patients is allowed in 1.5T horizontal bore scanners utilizing pulse sequences with reduced power. However, the use of 3T MRI in such patients is increasingly reported based on limited safety assessments. Here we present the results of comprehensive RF heating measurements for two commercially available DBS systems during MRI at 1.5T and 3T.PurposeTo assess the effect of imaging landmark, DBS lead configuration, and patient's body composition on RF heating of DBS leads during MRI at 1.5T and 3T.Study TypePhantom and ex vivo study.Population/Subjects/Phantom/Specimen/Animal ModelGel phantoms and cadaver brain.Field Strength/Sequence1.5T and 3T, T1‐weighted turbo spin echo.AssessmentRF heating was measured at the tips of DBS leads implanted in brain‐mimicking gel. Image artifact was assessed in a cadaver brain implanted with an isolated DBS lead.Statistical TestsDescriptive.ResultsWe observed substantial fluctuation in RF heating, mainly affected by phantom composition and DBS lead configuration, ranging from 0.14°C to 23.73°C at 1.5T, and from 0.10°C to 7.39°C at 3T. The presence of subcutaneous fat substantially altered RF heating at the electrode tips (3.06°C < ∆T < 19.05° C). Introducing concentric loops in the extracranial portion of the lead at the surgical burr hole reduced RF heating by up to 89% at 1.5T and up to 98% at 3T compared to worst‐case heating scenarios.Data ConclusionDevice configuration and patient's body composition substantially altered the RF heating of DBS leads during MRI. Interestingly, certain lead trajectories consistently reduced RF heating and image artifact.Level of Evidence 1Technical Efficacy Stage 1J. MAGN. RESON. IMAGING 2021;53:599–610.
Purpose
Patients with active implants such as deep brain stimulation (DBS) devices are often denied access to MRI due to safety concerns associated with the radiofrequency (RF) heating of their electrodes. The majority of studies on RF heating of conductive implants have been performed in horizontal close‐bore MRI scanners. Vertical MRI scanners which have a 90° rotated transmit coil generate fundamentally different electric and magnetic field distributions, yet very little is known about RF heating of implants in this class of scanners. We performed numerical simulations as well as phantom experiments to compare RF heating of DBS implants in a 1.2T vertical scanner (OASIS, Hitachi) compared to a 1.5T horizontal scanner (Aera, Siemens).
Methods
Simulations were performed on 90 lead models created from post‐operative CT images of patients with DBS implants. Experiments were performed with wires and commercial DBS devices implanted in an anthropomorphic phantom.
Results
We found significant reduction of 0.1 g‐averaged specific absorption rate (30‐fold, P < 1 × 10−5) and RF heating (9‐fold, P < .026) in the 1.2T vertical scanner compared to the 1.5T conventional scanner.
Conclusion
Vertical MRI scanners appear to generate lower RF heating around DBS leads, providing potentially heightened safety or the flexibility to use sequences with higher power levels than on conventional systems.
The majority of studies that assess magnetic resonance imaging (MRI) induced radiofrequency (RF) heating of the tissue when active electronic implants are present have been performed in horizontal, closed-bore MRI systems. Vertical, open-bore MRI systems have a 90° rotated magnet and a fundamentally different RF coil geometry, thus generating a substantially different RF field distribution inside the body. Little is known about the RF heating of elongated implants such as deep brain stimulation (DBS) devices in this class of scanners. Here, we conducted the first large-scale experimental study investigating whether RF heating was significantly different in a 1.2 T vertical field MRI scanner (Oasis, Fujifilm Healthcare) compared to a 1.5 T horizontal field MRI scanner (Aera, Siemens Healthineers). A commercial DBS device mimicking 30 realistic patient-derived lead trajectories extracted from postoperative computed tomography images of patients who underwent DBS surgery at our institution was implanted in a multi-material, anthropomorphic phantom. RF heating around the DBS lead was measured during four minutes of high-SAR RF exposure. Additionally, we performed electromagnetic simulations with leads of various internal structures to examine this effect on RF heating. When controlling for RMS B1+, the temperature increase around the DBS lead-tip was significantly lower in the vertical scanner compared to the horizontal scanner (0.33 ± 0.24°C vs. 4.19 ± 2.29°C). Electromagnetic simulations demonstrated up to a 17-fold reduction in the maximum of 0.1g-averaged SAR in the tissue surrounding the lead-tip in the vertical scanner compared to the horizontal scanner. Results were consistent across leads with straight and helical internal wires. Radiofrequency heating and power deposition around the DBS lead-tip were substantially lower in the 1.2 T vertical scanner compared to the 1.5 T horizontal scanner. Simulations with different lead structures suggest that the results may extend to leads from other manufacturers.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.