Purpose: To assess how the current practice of newborn screening (NBS) for homocystinurias compares with published recommendations. Methods: Twenty-two of 32 NBS programmes from 18 countries screened for at least one form of homocystinuria. Centres provided pseudonymised NBS data from patients with cystathionine beta-synthase deficiency (CBSD, n = 19), methionine adenosyltransferase I/III deficiency (MATI/IIID, n = 28), combined remethylation disorder (cRMD, n = 56) and isolated remethylation disorder (iRMD), including methylenetetrahydrofolate reductase deficiency (MTHFRD) (n = 8). Markers and decision limits were converted to multiples of the median (MoM) to allow comparison between centres.
Due to the favourable outcome of early treated patients, NBS for homocystinurias is recommended. To improve NBS, decision limits should be revised considering the population median. Relevant markers should be combined; use of the postanalytical tools offered by the CLIR project (Collaborative Laboratory Integrated Reports, which considers, e.g. birth weight and gestational age) is recommended. tHcy and methylmalonic acid should be implemented as second-tier markers.
Background: Blood samples for neonatal screening for inborn errors of metabolism are collected and shipped on standardized filter paper cards. Occasionally these samples are contaminated with EDTA, which is often used for anticoagulation. EDTA may interfere with newborn screening tests based on lanthanide fluorescence and thus lead to false-negative or false-positive results.
Methods: We used tandem mass spectrometry (MS/MS) to detect EDTA in dried blood spots by use of an extra experiment that was integrated into the standard MS/MS neonatal screening and did not require an additional sample spot, nor extra time or work. We analyzed the influence of different blood sampling procedures on lanthanide fluorescence tests for thyroid-stimulating hormone (TSH) and 17-hydroxyprogesterone (17-OHP).
Results: EDTA was increased in 138 of 190 000 newborn screening samples, 27 of which caused false- positive results in the immunoassay for 17-OHP. No false-negative TSH results were found. False-positive results in the 17-OHP test occurred when EDTA concentrations were >2.0 g/L; the TSH test, however, produced false negatives only when EDTA concentrations were >3.0 g/L. Using EDTA-containing devices the procedure of blood collection significantly influenced the concentration of the anticoagulant.
Conclusion: Addition of EDTA quantification into standard MS/MS tests is a simple and useful method to avoid false-positive or false-negative neonatal screening results in lanthanide fluorescence–based tests.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.