Objectives
The aims of this study were to evaluate the agreement between cone-beam computed tomography (CBCT) and multislice computed tomography for the characterization of extremity fractures and to compare image quality, radiation dose, and patient tolerance.
Methods
Thirty-six patients with suspected fracture affecting distal extremities or who required preoperative fracture assessment were enrolled prospectively. Each patient underwent CBCT and multislice computed tomography the same day. Both examinations were evaluated independently twice by 2 trained radiologists using the Müller AO classification for fracture characterization.
Results
Cohen κ coefficient for agreement between the imaging techniques was almost perfect for fracture characterization, κ = 0.94 [95% confidence interval, 0.91–0.98]. There was substantial to almost perfect agreement for secondary findings. Cone-beam computed tomography was well tolerated and significantly less irradiant and had better subjective image.
Conclusions
An excellent agreement between both imaging techniques was found. This confirms the ability of CBCT to assess fractures and its potential in the management of patients with distal limb trauma.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.