The MR-Sim setup and automatic sCT generation methods using standard MR sequences generates realistic contours and electron densities for prostate cancer radiation therapy dose planning and digitally reconstructed radiograph generation.
An novel automatic framework for gold fiducial marker detection in MRI is proposed and evaluated with detection accuracies comparable to manual detection. When radiation therapists are unable to determine the seed location in MRI, they refer back to the planning CT (only available in the existing clinical framework); similarly, an automatic quality control is built into the automatic software to ensure that all gold seeds are either correctly detected or a warning is raised for further manual intervention.
To clinically implement MRI simulation or MRI-alone treatment planning requires comprehensive end-to-end testing to ensure an accurate process. The purpose of this study was to design and build a geometric phantom simulating a human male pelvis that is suitable for both CT and MRI scanning and use it to test geometric and dosimetric aspects of MRI simulation including treatment planning and digitally reconstructed radiograph (DRR) generation.A liquid filled pelvic shaped phantom with simulated pelvic organs was scanned in a 3T MRI simulator with dedicated radiotherapy couch-top, laser bridge and pelvic coil mounts. A second phantom with the same external shape but with an internal distortion grid was used to quantify the distortion of the MR image. Both phantoms were also CT scanned as the gold-standard for both geometry and dosimetry. Deformable image registration was used to quantify the MR distortion. Dose comparison was made using a seven-field IMRT plan developed on the CT scan with the fluences copied to the MR image and recalculated using bulk electron densities. Without correction the maximum distortion of the MR compared with the CT scan was 7.5 mm across the pelvis, while this was reduced to 2.6 and 1.7 mm by the vendor's 2D and 3D correction algorithms, respectively. Within the locations of the internal organs of interest, the distortion was <1.5 and <1 mm with 2D and 3D correction algorithms, respectively. The dose at the prostate isocentre calculated on CT and MRI images differed by 0.01% (1.1 cGy). Positioning shifts were within 1 mm when setup was performed using MRI generated DRRs compared to setup using CT DRRs.The MRI pelvic phantom allows end-to-end testing of the MRI simulation workflow with comparison to the gold-standard CT based process. MRI simulation was found to be geometrically accurate with organ dimensions, dose distributions and DRR based setup within acceptable limits compared to CT.
Routine quality assurance (QA) is necessary and essential to ensure MR scanner performance. This includes geometric distortion, slice positioning and thickness accuracy, high contrast spatial resolution, intensity uniformity, ghosting artefact and low contrast object detectability. However, this manual process can be very time consuming. This paper describes the development and validation of an open source tool to automate the MR QA process, which aims to increase physicist efficiency, and improve the consistency of QA results by reducing human error. The OSAQA software was developed in Matlab and the source code is available for download from http://jidisun.wix.com/osaqa-project/. During program execution QA results are logged for immediate review and are also exported to a spreadsheet for long-term machine performance reporting. For the automatic contrast QA test, a user specific contrast evaluation was designed to improve accuracy for individuals on different display monitors. American College of Radiology QA images were acquired over a period of 2 months to compare manual QA and the results from the proposed OSAQA software. OSAQA was found to significantly reduce the QA time from approximately 45 to 2 min. Both the manual and OSAQA results were found to agree with regard to the recommended criteria and the differences were insignificant compared to the criteria. The intensity homogeneity filter is necessary to obtain an image with acceptable quality and at the same time keeps the high contrast spatial resolution within the recommended criterion. The OSAQA tool has been validated on scanners with different field strengths and manufacturers. A number of suggestions have been made to improve both the phantom design and QA protocol in the future.
The purpose of this study was to investigate performance of the couch and coil mounts designed for MR‐simulation prostate scanning using data from ten volunteers. Volunteers were scanned using the standard MR scanning protocol with the MR coil directly strapped on the external body and the volunteer lying on the original scanner table. They also were scanned using a MR‐simulation table top and pelvic coil mounts. MR images from both setups were compared in terms of body contour variation and image quality effects within particular organs of interest. Six‐field conformal plans were generated on the two images with assigned bulk density for dose calculation. With the MR‐simulation devices, the anterior skin deformation was reduced by up to 1.7 cm. The hard tabletop minimizes the posterior body deformation which can be up to 2.3 cm on the standard table, depending on the weight of volunteer. The image signal‐to‐noise ratio reduced by 14% and 25% on large field of view (FOV) and small FOV images, respectively, after using the coil mount; the prostate volume contoured on two images showed difference of 1.05±0.66 cm3. The external body deformation caused a mean dose reduction of 0.6±0.3 Gy, while the coverage reduced by 22%±13% and 27%±6% in V98 and V100, respectively. A dedicated MR simulation setup for prostate radiotherapy is essential to ensure the agreement between planning anatomy and treatment anatomy. The image signal was reduced after applying the coil mount, but no significant effect was found on prostate contouring.PACS numbers: 87.55.D‐, 87.61.‐c, 87.57.C‐
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.
customersupport@researchsolutions.com
10624 S. Eastern Ave., Ste. A-614
Henderson, NV 89052, USA
This site is protected by reCAPTCHA and the Google Privacy Policy and Terms of Service apply.
Copyright © 2024 scite LLC. All rights reserved.
Made with 💙 for researchers
Part of the Research Solutions Family.