Introduction Patients on intravenous heparin require regular activated partial thromboplastin time monitoring. Laboratory-based activated partial thromboplastin time assays necessitate a delay between blood sampling and dose adjustment. Point-of-care testing could permit immediate dose adjustments, potentially enabling tighter control of anticoagulation. Aim To assess equivalence of activated partial thromboplastin time measured by conventional laboratory assay and by a novel proprietary point-of-care testing system (Hemochron Response, ITC, Thoratec Corporation, Edison, NJ) among surgical ward patients on intravenous heparin. Methods A total of 39 blood samples from patients on intravenous heparin were tested with both laboratory and point-of-care assays. Assay equivalence was assessed by Bland-Altman analysis. Results. Point-of-care measurements exceeded laboratory activated partial thromboplastin time by a mean of 15 seconds (standard deviation 19). In 19 cases (49%), the point-of-care measurement would have resulted in different heparin dosing from the laboratory activated partial thromboplastin time. Conclusions The Hemochron Response system is not sufficiently accurate for routine ward use compared with laboratory activated partial thromboplastin time assays.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.