Background
Treating severe forms of the acute respiratory distress syndrome and cardiac failure, extracorporeal membrane oxygenation (ECMO) has become an established therapeutic option. Neonatal or pediatric patients receiving ECMO, and patients undergoing extracorporeal CO2 removal (ECCO2R) represent low-flow applications of the technology, requiring lower blood flow than conventional ECMO. Centrifugal blood pumps as a core element of modern ECMO therapy present favorable operating characteristics in the high blood flow range (4 L/min–8 L/min). However, during low-flow applications in the range of 0.5 L/min–2 L/min, adverse events such as increased hemolysis, platelet activation and bleeding complications are reported frequently.
Methods
In this study, the hemolysis of the centrifugal pump DP3 is evaluated both in vitro and in silico, comparing the low-flow operation at 1 L/min to the high-flow operation at 4 L/min.
Results
Increased hemolysis occurs at low-flow, both in vitro and in silico. The in-vitro experiments present a sixfold higher relative increased hemolysis at low-flow. Compared to high-flow operation, a more than 3.5-fold increase in blood recirculation within the pump head can be observed in the low-flow range in silico.
Conclusions
This study highlights the underappreciated hemolysis in centrifugal pumps within the low-flow range, i.e. during pediatric ECMO or ECCO2R treatment. The in-vitro results of hemolysis and the in-silico computational fluid dynamic simulations of flow paths within the pumps raise awareness about blood damage that occurs when using centrifugal pumps at low-flow operating points. These findings underline the urgent need for a specific pump optimized for low-flow treatment. Due to the inherent problems of available centrifugal pumps in the low-flow range, clinicians should use the current centrifugal pumps with caution, alternatively other pumping principles such as positive displacement pumps may be discussed in the future.
In the development of new hemocompatible biomaterials, surface modification appears to be a suitable method in order to reduce the thrombogenetic potential of such materials. In this study, polycarbonate-urethane (PCU) tubes with different surface microstructures to be used for aortic heart valve models were investigated with regard to the thrombogenicity. The surface structures were produced by using a centrifugal casting process for manufacturing PCU tubes with defined casting mold surfaces which are conferred to the PCU surface during the process. Tubes with different structures defined by altering groove widths were cut into films and investigated under dynamic flow conditions in contact with porcine blood. The analysis was carried out by laser scanning microscopy which allowed for counting various morphological types of platelets with regard to the grade of activation. The comparison between plain and shaped PCU samples showed that the surface topography led to a decline of the activation of the coagulation cascade and thus to the reduction of the fibrin synthesis. Comparing different types of structures revealed that smooth structures with a small groove width (d ~ 3 μm) showed less platelet activation as well as less adhesion in contrast to a distinct wave structure (d ~ 90 μm). These results prove surface modification of polymer biomaterials to be a suitable method for reducing thrombogenicity and hence give reason for further alterations and improvements.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.