Purpose
A new acquisition and reconstruction method called T2 Shuffling is presented for volumetric fast spin-echo (3D FSE) imaging. T2 Shuffling reduces blurring and recovers many images at multiple T2 contrasts from a single acquisition at clinically feasible scan times (6 to 7 minutes).
Theory and Methods
The parallel imaging forward model is modified to account for temporal signal relaxation during the echo train. Scan efficiency is improved by acquiring data during the transient signal decay and by increasing echo train lengths without loss in SNR. By (1) randomly shuffling the phase encode view ordering, (2) constraining the temporal signal evolution to a low-dimensional subspace, and (3) promoting spatio-temporal correlations through locally low rank regularization, a time series of virtual echo time images is recovered from a single scan. A convex formulation is presented that is robust to partial voluming and RF field inhomogeneity.
Results
Retrospective under-sampling and in vivo scans confirm the increase in sharpness afforded by T2 Shuffling. Multiple image contrasts are recovered and used to highlight pathology in pediatric patients. A proof-of-principle method is integrated into a clinical musculoskeletal imaging workflow.
Conclusion
The proposed T2 Shuffling method improves the diagnostic utility of 3D FSE by reducing blurring and producing multiple image contrasts from a single scan.
Purpose
To develop and evaluate motion-compensation and compressed-sensing techniques in 4D flow MRI for anatomical assessment in a comprehensive ferumoxytol-enhanced congenital heart disease (CHD) exam.
Materials and Methods
A Cartesian 4D flow sequence was developed to enable intrinsic navigation and two sampling schemes: VDPoisson and VDRad. Four compressed-sensing methods were developed: A) VDPoisson scan reconstructed using spatial wavelets, B) added temporal total variation to A, C) VDRad scan using the same reconstruction as in B, and D) added motion compensation to C.
With IRB approval and HIPAA compliance, 23 consecutive patients (8 females, mean 6.3 years) referred for ferumoxytol-enhanced CHD 3T MRI were recruited. Images were acquired and reconstructed using methods A–D. Two cardiovascular radiologists independently scored the images on a 5-point scale. These readers performed a paired wall motion and functional assessment between method D and 2D bSSFP CINE for 16 cases.
Results
Method D had higher diagnostic image quality for most anatomical features (mean 3.8–4.8) compared to A (2.0–3.6), B (2.2–3.7), and C (2.9–3.9) with P < 0.05 with good inter-observer agreement (κ ≥ 0.49). Method D had similar or better assessment of myocardial borders and cardiac motion compared to 2D bSSFP (P < 0.05, κ ≥ 0.77). All methods had good internal agreement in comparing aortic with pulmonic flow (BA mean < 0.02%, r > 0.85) and compared to method A (BA mean < 0.13%, r > 0.84) with P < 0.01.
Conclusion
Flow, functional, and anatomical assessment in CHD with ferumoxytol-enhanced 4D flow is feasible and can be significantly improved using motion compensation and compressed sensing.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.