An amperometric detector for hydrophobic ions based on a plasticized poly(vinyl) chloride (PVC) membrane incorporated in a flow-injection system was developed. A four-electrode potentiostat with ohmic drop compensation was used, while a flow-through cell incorporated the four electrodes and the membrane, which contained tetrabutylammonium tetraphenylborate. When the influence of the applied potential and of the flow-injection variables on the determination of tetrabutylammonium was studied, a linear relationship was observed between current peak height and ion concentration over a range of 5 Â 10 À6 ± 6 Â 10 À5 M tetrabutylammonium. Good repeatability and between-day reproducibility and high sample frequency were obtained. The effect of other ions was studied. Two different amperometric methods, indirect and direct, were also developed for the determination of dodecylsulfate in the concentration range 3 Â 10 À5 ± 9 Â 10 À4 M.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.