Osteogenesis Imperfecta (IO) is a bone disease mainly characterized by the low bone density that produces common fractures in children around 0–7 years. The use of metal implants is a typical treatment of this disease. The intramedullary telescopic nail (ITN) was inspired by the progressive growth in the long bones such as the femur or humerus during children’s aging. This work shows an experimental assessment of the ITN’s, focusing on their fixation; the proposed improvements in the design of the intramedullary nail studied include the separation of the element into two parts for telescopic enlargement, minimal invasive fixation through the distal anchorage, and the double auto-drilled end for fixation on the distal and proximal section of the bone. The samples were manufactured in 316 L steel and mounted on specialized jaws to replicate the implants’ boundary conditions. The experimental test was repeated three times to report the intramedullary telescopic nail’s behavior at three lengths. The results show that the device supports only 79.06 N when not at extension length. However, if the device is extended 150% it will support 46.87 N which suggests that intramedullary telescopic nails can only increase by 25% of their original length before they fail.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.