We describe a human and large animal Langendorff experimental apparatus for live electrophysiological studies and measure the electrophysiological changes due to gap junction uncoupling in human and porcine hearts. The resultant ex vivo intact human and porcine model can bridge the translational gap between smaller simple laboratory models and clinical research. In particular, electrophysiological models would benefit from the greater myocardial mass of a large heart due to its effects on far-field signal, electrode contact issues and motion artefacts, consequently more closely mimicking the clinical setting. Porcine (n = 9) and human (n = 4) donor hearts were perfused on a custom-designed Langendorff apparatus. Epicardial electrograms were collected at 16 sites across the left atrium and left ventricle. A total of 1 mM of carbenoxolone was administered at 5 ml/min to induce cellular uncoupling, and then recordings were repeated at the same sites. Changes in electrogram characteristics were analysed. We demonstrate the viability of a controlled ex vivo model of intact porcine and human hearts for electrophysiology with pharmacological modulation. Carbenoxolone reduces cellular coupling and changes contact electrogram features. The time from stimulus artefact to (-dV/dt)max increased between baseline and carbenoxolone (47.9 ± 4.1–67.2 ± 2.7 ms) indicating conduction slowing. The features with the largest percentage change between baseline and carbenoxolone were fractionation + 185.3%, endpoint amplitude − 106.9%, S-endpoint gradient + 54.9%, S point − 39.4%, RS ratio + 38.6% and (-dV/dt)max − 20.9%. The physiological relevance of this methodological tool is that it provides a model to further investigate pharmacologically induced pro-arrhythmic substrates.
Aims
The response to high frequency stimulation (HFS) is used to locate putative sites of ganglionated plexuses (GPs), which are implicated in triggering atrial fibrillation (AF). To identify topological and immunohistochemical characteristics of presumed GP sites functionally identified by HFS.
Methods and results
Sixty-three atrial sites were tested with HFS in four Langendorff-perfused porcine hearts. A 3.5 mm tip quadripolar ablation catheter was used to stimulate and deliver HFS to the left and right atrial epicardium, within the local atrial refractory period. Tissue samples from sites triggering atrial ectopy/AF (ET) sites and non-ET sites were stained with choline acetyltransferase (ChAT) and tyrosine hydroxylase (TH), for quantification of parasympathetic and sympathetic nerves, respectively. The average cross-sectional area (CSA) of nerves was also calculated. Histomorphometry of six ET sites (9.5%) identified by HFS evoking at least a single atrial ectopic was compared with non-ET sites. All ET sites contained ChAT-immunoreactive (ChAT-IR) and/or TH-immunoreactive nerves (TH-IR). Nerve density was greater in ET sites compared to non-ET sites (nerves/cm2: 162.3 ± 110.9 vs. 69.65 ± 72.48; P = 0.047). Overall, TH-IR nerves had a larger CSA than ChAT-IR nerves (µm2: 11 196 ± 35 141 vs. 2070 ± 5841; P < 0.0001), but in ET sites, TH-IR nerves were smaller than in non-ET sites (µm2: 6021 ± 14 586 vs. 25 254 ± 61 499; P < 0.001).
Conclusions
ET sites identified by HFS contained a higher density of smaller nerves than non-ET sites. The majority of these nerves were within the atrial myocardium. This has important clinical implications for devising an effective therapeutic strategy for targeting autonomic triggers of AF.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.