Totally implantable venous access devices (TIVADs) are frequently used for the administration of chemotherapy or parenteral nutrition and have proved to be safe and improve patient quality of life during treatment. Catheter-related infections are the most common complication, whereas catheter fracture and embolization is rarely seen. We present a case of a 61-year-old cancer patient in which the TIVAD fractured and embolized to the patient’s left knee 18 months after its initial placement. An endovascular procedure with intraoperative imaging revealed that the catheter had embolized to the popliteal artery and was successfully removed, the only explanation being the presence of an atrial septal defect. A bubble echocardiogram was unsuccessful in confirming the diagnosis. The catheter fracture could be related to an intravascular procedure that was performed 18 months prior to remove fibrin sheaths as a cause of port malfunction, or it could be a case of pinch-off syndrome.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.