The goal of this study was to exploit the excellent spatial resolution characteristics of a position-sensitive silicon photomultiplier (SiPM) and develop a high-resolution depth-of-interaction (DOI) encoding positron emission tomography (PET) detector module. The detector consists of a 30 × 30 array of 0.445 × 0.445 × 20 mm polished LYSO crystals coupled to two 15.5 × 15.5 mm linearly-graded SiPM (LG-SiPM) arrays at both ends. The flood histograms show that all the crystals in the LYSO array can be resolved. The energy resolution, the coincidence timing resolution and the DOI resolution were 21.8 ± 5.8%, 1.23 ± 0.10 ns and 3.8 ± 1.2 mm, respectively, at a temperature of -10 °C and a bias voltage of 35.0 V. The performance did not degrade significantly for event rates of up to 130 000 counts s. This detector represents an attractive option for small-bore PET scanner designs that simultaneously emphasize high spatial resolution and high detection efficiency, important, for example, in preclinical imaging of the rodent brain with neuroreceptor ligands.
In this work, a small animal PET scanner named SIAT aPET was developed using dual-ended readout depth encoding detectors to simultaneously achieve high spatial resolution and high sensitivity. The scanner consists of four detector rings with 12 detector modules per ring; the ring diameter is 111 mm and the axial field of view (FOV) is 105.6 mm. The images are reconstructed using an ordered subset expectation maximization (OSEM) algorithm. The spatial resolution of the scanner was measured by using a 22Na point source at the center axial FOV with different radial offsets. The sensitivity of the scanner was measured at center axis of the scanner with different axial positions. The count rate performance of the system was evaluated by scanning mouse-sized and rat-sized phantoms. An ultra-micro hot-rods phantom and two mice injected with 18F-NaF and 18F-FDG were scanned on the scanner. An average depth of interaction (DOI) resolution of 1.96 mm, energy resolution of 19.1% and timing resolution of 1.20 ns were obtained for the detector. Average spatial resolutions of 0.82 mm and 1.16 mm were obtained up to a distance of 30 mm radially from the center of the FOV when reconstructing a point source in 1% and 10% warm backgrounds, respectively, using OSEM reconstruction with 16 subsets and 10 iterations. Sensitivities of 16.0% and 11.9% were achieved at center of the scanner for energy windows of 250–750 keV and 350–750 keV respectively. Peak noise equivalent count rates (NECRs) of 324 kcps and 144 kcps were obtained at an activity of 26.4 MBq for the mouse-sized and rat-sized phantoms. Rods of 1.0 mm diameter can be visually resolved from the image of the ultra-micro hot-rods phantom. The capability of the scanner was demonstrated by high quality in-vivo mouse images.
The performance of an 8 × 8 array of 6.0 × 6.0 mm2 (active area) SiPMs was evaluated for PET applications using crystal arrays with different pitch sizes (3.4 mm, 1.5 mm, 1.35 mm and 1.2 mm) and custom designed five-channel front-end readout electronics (four channels for position information and one channel for timing information). The total area of this SiPM array is 57.4 × 57.4 mm2, and the pitch size is 7.2 mm. It was fabricated using enhanced blue sensitivity SiPMs (MicroFB-60035-SMT) with peak spectral sensitivity at 420 nm. The performance of the SiPM array was characterized by measuring flood histogram decoding quality, energy resolution, timing resolution and saturation at several bias voltages (from 25.0 V to 30.0 V in 0.5 V intervals) and two different temperatures (5 °C and 20 °C). Results show that the best flood histogram was obtained at a bias voltage of 28.0 V and 5 °C and an array of polished LSO crystals with a pitch as small as 1.2 mm can be resolved. No saturation was observed up to a bias voltage of 29.5 V during the experiments, due to adequate light sharing between SiPMs. Energy resolution and timing resolution at 5 °C ranged from 12.7 ± 0.8% to 14.6 ± 1.4 % and 1.58 ± 0.13 ns to 2.50 ± 0.44 ns, for crystal array pitch sizes of 3.4 mm and 1.2 mm respectively. Superior flood histogram quality, energy resolution and timing resolution were obtained with larger crystal array pitch sizes and at lower temperature. Based on our findings, we conclude that this large-area SiPM array can serve as a suitable photodetector for high-resolution small-animal PET or dedicated human brain PET scanners.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.
customersupport@researchsolutions.com
10624 S. Eastern Ave., Ste. A-614
Henderson, NV 89052, USA
This site is protected by reCAPTCHA and the Google Privacy Policy and Terms of Service apply.
Copyright © 2024 scite LLC. All rights reserved.
Made with 💙 for researchers
Part of the Research Solutions Family.