The article presents spectral sensitivity curves of 22 typical luxmeters and 3 smartphones as a result of laboratory measurements. Based on the measurement data, errors in the spectral correction of the considered instruments and devices were calculated. Due to the fact that the value of the measured illuminance is influenced by the spectral characteristics of the radiation, in order to determine the spectral corrections coefficients, 10 light sources with different spectral distributions of radiation were taken into account. Spectral corrections factors were determined using two methods for all luxmeters and smartphones. This paper presents the relationship between the shape of the spectral sensitivity curves of a given luxmeter and the measured values of f’1 errors. Moreover, the limitations related to the use of spectral corrections in environmental measurements with the use of a luxmeter were demonstrated.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.