The objective of this study was to determine the forces on dental prosthetic magnet keepers, with a view to assessing the potential for patient injury during magnetic resonance imaging (MRI). Four pre-formed keepers and one castable keeper alloy were tested. Magnetizations and high field susceptibilities were determined for each of the five specimens using data from a vibrating sample magnetometer. The magnetic field intensity with respect to distance from the main magnet coil was obtained from the manufacturer (1-5 tesla General Electric Signa Imaging System). A plot of force versus distance from the main coil and the maximum force at the magnet portal was determined for each specimen. The maximum forces ranged from 0.12-0.24 N for the pre-formed keepers and 3.67 MNm-3 for the castable alloy. It was concluded that the risk of patient injury by displacement is minimal, if the keepers are properly attached to supporting structures.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.